Since temperature affects the measuring accuracy of a pressure sensor, there always remains a little temperature error in the rated temperature range despite a variety of compensation measures. Attack is often expressed in the info sheets of manufacturers of pressure sensors as temperature coefficient (abbr. TC). This coefficient describes a (linear) error, starting from a reference point, which generally is room temperature.
Accordingly, the temperature error at room temperature is zero and increases with increasing difference of the temperature from room temperature with the specified coefficient in linear fashion (see figure). Basically, not only the zero-point temperature coefficient (abbr. TC0), but additionally that of the span (TCC) should be considered individually. The zero-point error describes the effect of temperature on the zero signal. The error of the span specifies the effect of temperature on the entire scale value. Kill of zero point and of span are often specified as magnitudes, and therefore they could be either positive or negative.
If in an individual instrument the zero-point error has the same sign as the error of the span, these two temperature errors may add up in worst case. A typical value for the common zero-point temperature coefficient of a pressure sensor is: 0.2 % / 10 K. Link to WIKA pressure sensors.